EPSRC National Crystallography Service

Data Collection Summary

Summary report for Directory: diska/02src179

Report generated Apr 27, 2002; 12:00:16

Unit cell

50787 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement
 
 
Symmetry used
in scalepack
p1
a (Angstrom) 14.3182 +/- 0.0002
b (Angstrom) 15.9481 +/- 0.0002
c (Angstrom) 19.1468 +/- 0.0004
alpha (°) 96.1370 +/- 0.0006
beta (°) 108.2309 +/- 0.0007
gamma (°) 113.5006 +/- 0.0010
Volume (A**3) 3671.07 +/- 0.11
Mosaicity (°) 0.7260 +/- 0.0010

Data collection

Summary

Total number of images collected 231
Total exposure time 12.4 hours
Data collection exposure time 12.4 hours
Data collection wall-clock time 12.9 hours

Experimental Conditions

Wavelength 0.71073 A
Generator setting 50kV 75mA
Crystal to detector distance 30.00 mm

Scans

Type Name # images Total
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collection s01f 181 362.0° phi 2.000° 200 seconds Yes
data collection s02f 16  32.0° omega 2.000° 200 seconds Yes
data collection s03f 26  52.0° omega 2.000° 200 seconds Yes
Phi/Chi i01f - i08f 8 20 seconds

Scalepack Scaling

Deleted observations

Zero sigma or profile test    2
Overload or incomplete profile 1429
Sigma cutoff  333
High resolution limit  184

Final Data Set

Scale factor 10.00
Number of 'full' reflections  11165
Number of 'partial' reflections  83038
Total number of integrated reflections  56364
Total number of unique reflections  16498
Data completeness   98.4%
Resolution range 7.00-0.77 A
Theta range 2.91°-27.48°
Average Intensity    62.3
Average Sigma(I)     3.0
Overall R-merge (linear)   0.054

Crystal Information

Collection Temperature   120 K
Crystal Size  0.26 x 0.12 x 0.03
Crystal Colour  colourless
Crystal Shape  plate

 
 

SORTAV Absorption correction

Max Transmission Factor
0.99489 
Min Transmission Factor
0.94699 

N.B. The scaling summary is redundant as outliers are treated during the absorption correction using SORTAV. The quoted data completeness is for the stated resolution ranges, and the Overall R-merge is that before the absorption correction. The SORTAV transmission factors are based on a crude approximation and the expected formula (not always correct!). For more information visit the service web site at: http://www.soton.ac.uk/~xservice/