EPSRC National Crystallography Service

Data Collection Summary

Summary report for Directory: diska/00src612

Report generated Dec 19, 2000; 12:02:18

Unit cell

5079 reflections with 2.91°<theta<39.39° (resolution between 7.00A and 0.56A) were used for unit cell refinement

Symmetry used
in scalepack
p1
a (Angstrom)9.16150 +/- 0.00010
b (Angstrom)10.1143 +/- 0.0002
c (Angstrom)18.1371 +/- 0.0005
alpha (°)88.5860 +/- 0.0009
beta (°)80.2376 +/- 0.0009
gamma (°)70.9954 +/- 0.0016
Volume (A**3)1565.16 +/- 0.06
Mosaicity (°)0.342 +/- 0.002

Data collection

Summary

Total number of images collected144
Total exposure time47.2 minutes
Data collection exposure time45.8 minutes
Data collection wall-clock time63.2 minutes

Experimental Conditions

Wavelength0.71073 A
Generator setting50.00kV 90.00mA
Crystal to detector distance40.00 mm

Scans

TypeName# imagesTotal
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collections01f91182.0° phi2.000°20 secondsYes
data collections02f22 44.0° omega2.000°20 secondsYes
data collections03f23 46.0° omega2.000°20 secondsYes

Scaling

Deleted observations

Overload or incomplete profile 523
Sigma cutoff   1

Final Data Set

Scale factor10.00
Number of 'full' reflections  8128
Number of 'partial' reflections  6041
Total number of integrated reflections 10586
Total number of unique reflections  6323
Data completeness  34.1%
Resolution range7.00-0.56 A
Theta range2.91°-39.39°
Average Intensity  145.3
Average Sigma(I)    6.9
Overall R-merge (linear)  0.039

Crystal Information

Collection Temperature   150 K
Crystal Size    x   x  
Crystal Colour  
Crystal Shape  

SORTAV Absorption correction Summary

R(merge) on F2
Before
After
R
0.053
0.048
Rw
7.817
7.914
Max Transmission Factor
0.976E+00
Min Transmission Factor
0.858E+00

N.B. The scaling summary is redundant as outliers are treated during the absorption correction using SORTAV. The quoted data completeness is for the stated resolution ranges, and the Overall R-merge is that before the absorption correction. The SORTAV transmission factors are based on a crude approximation and the expected formula (not always correct!). For more information visit the service web site at: http://www.soton.ac.uk/~xservice/