EPSRC National Crystallography Service

Data Collection Summary

Summary report for Directory: diska/01src685

Report generated Jul 17, 2001; 14:04:14

Unit cell

13451 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement
 
 
Symmetry used
in scalepack
p1
a (Angstrom) 6.9951 +/- 0.0004
b (Angstrom) 9.7142 +/- 0.0005
c (Angstrom) 16.2552 +/- 0.0011
alpha (°) 78.602 +/- 0.002
beta (°) 85.273 +/- 0.002
gamma (°) 82.743 +/- 0.003
Volume (A**3) 1072.24 +/- 0.11
Mosaicity (°) 0.819 +/- 0.004

Data collection

Summary

Total number of images collected 224
Total exposure time 146.1 minutes
Data collection exposure time 144.8 minutes
Data collection wall-clock time 175.2 minutes

Experimental Conditions

Wavelength 0.71073 A
Generator setting 50kV 50mA
Crystal to detector distance 30.00 mm

Scans

Type Name # images Total
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collection s01f 179 358.0° phi 2.000° 40 seconds Yes
data collection s02f 16  32.0° omega 2.000° 40 seconds Yes
data collection s03f 21  42.0° omega 2.000° 40 seconds Yes
Phi/Chi i01f - i08f 8 10 seconds

Scalepack Scaling

Deleted observations

Zero sigma or profile test   10
Overload or incomplete profile  358
Sigma cutoff   29

Final Data Set

Scale factor 10.00
Number of 'full' reflections   2664
Number of 'partial' reflections  25051
Total number of integrated reflections  15957
Total number of unique reflections   4787
Data completeness   96.6%
Resolution range 7.00-0.77 A
Theta range 2.91°-27.48°
Average Intensity    11.8
Average Sigma(I)     1.6
Overall R-merge (linear)   0.172

Crystal Information

Collection Temperature   120 K
Crystal Size  0.24 x 0.06 x 0.04
Crystal Colour  colourless
Crystal Shape  rod

SORTAV Absorption correction Summary

R(merge) on F2
Before
After
R
0.141 
0.139 
Rw
1.783 
1.788 
Max Transmission Factor
0.799. A( 
Min Transmission Factor
0.799. A( 

N.B. The scaling summary is redundant as outliers are treated during the absorption correction using SORTAV. The quoted data completeness is for the stated resolution ranges, and the Overall R-merge is that before the absorption correction. The SORTAV transmission factors are based on a crude approximation and the expected formula (not always correct!). For more information visit the service web site at: http://www.soton.ac.uk/~xservice/