EPSRC National Crystallography Service

Data Collection Summary

Summary report for Directory: diska/01src590

Report generated Jun 27, 2001; 15:14:32

Unit cell

4106 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement

Symmetry used
in scalepack
p222
a (Angstrom)6.5149 +/- 0.0002
b (Angstrom)13.0381 +/- 0.0005
c (Angstrom)16.4355 +/- 0.0006
alpha (°) 90.000
beta (°) 90.000
gamma (°) 90.000
Volume (A**3)1396.06 +/- 0.09
Mosaicity (°)0.465 +/- 0.003

Data collection

Summary

Total number of images collected77
Total exposure time24.6 minutes
Data collection exposure time23.3 minutes
Data collection wall-clock time30.9 minutes

Experimental Conditions

Wavelength0.71073 A
Generator setting50kV 85mA
Crystal to detector distance30.00 mm

Scans

TypeName# imagesTotal
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collections01f69138.0° phi2.000°20 secondsYes
Phi/Chii01f - i08f810 seconds

Scalepack Scaling

Deleted observations

Overload or incomplete profile 109
Sigma cutoff   3
High resolution limit   2

Final Data Set

Scale factor10.00
Number of 'full' reflections  1064
Number of 'partial' reflections  8613
Total number of integrated reflections  6882
Total number of unique reflections  1846
Data completeness  98.5%
Resolution range7.00-0.77 A
Theta range2.91°-27.48°
Average Intensity   92.4
Average Sigma(I)    3.6
Overall R-merge (linear)  0.104

Crystal Information

Collection Temperature   120 K
Crystal Size    x   x  
Crystal Colour  
Crystal Shape  

SORTAV Absorption correction Summary

R(merge) on F2
Before
After
R
0.115
0.090
Rw
1.138
1.149
Max Transmission Factor
0.996E+00
Min Transmission Factor
0.627E+00

N.B. The scaling summary is redundant as outliers are treated during the absorption correction using SORTAV. The quoted data completeness is for the stated resolution ranges, and the Overall R-merge is that before the absorption correction. The SORTAV transmission factors are based on a crude approximation and the expected formula (not always correct!). For more information visit the service web site at: http://www.soton.ac.uk/~xservice/