Unit cell

32968 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement

Symmetry used
in scalepack
p2
a (Angstrom)7.9738 +/- 0.0007
b (Angstrom)8.4697 +/- 0.0009
c (Angstrom)10.8836 +/- 0.0011
alpha (°) 90.000
beta (°)102.652 +/- 0.006
gamma (°) 90.000
Volume (A**3)717.19 +/- 0.12
Mosaicity (°)2.141 +/- 0.004

Data collection

Summary

Total number of images collected423
Total exposure time69.3 minutes
Data collection exposure time67.9 minutes
Data collection wall-clock time102.2 minutes

Experimental Conditions

Wavelength0.71073 A
Crystal to detector distance35.00 mm

Scans

TypeName# imagesTotal
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collections01f226271.2° phi1.200°9 secondsYes
data collections02f97116.4° omega1.200°9 secondsYes
data collections03f92110.4° omega1.200°9 secondsYes
Phi/Chii01f - i08f810 seconds

Scalepack Scaling

Deleted observations

Rejected  20
Zero sigma or profile test  15
Overload or incomplete profile 528
Sigma cutoff  16
High resolution limit  10

Final Data Set

Scale factor10.00
Number of 'full' reflections     0
Number of 'partial' reflections 10793
Total number of integrated reflections 10071
Total number of unique reflections  1745
Data completeness  99.9%
Resolution range7.00-0.77 A
Theta range2.91°-27.48°
Average Intensity   56.6
Average Sigma(I)    2.4
Overall R-merge (linear)  0.078

Sadabs Results

Estimated minimum and maximum transmission: 0.5844 0.7456

Metadata

  Group    Service  
  Operator   Sam Callear  
  Sample Owner   Prof P.G Taylor  
  Local Code   SM 119bis  
  Formula   C15H11Cl3F6N2O2Si  
  Crystal Colour    Colourless  
  Crystal Habit    Block  
  Crystal Size   0.33 x 0.23 x 0.17 (mm3) 
  Temperature   120(K) 
  Generator   50 (kV)   85 (mA) 
  Primary Solvent     
  Other Solvents     

File Explanations

SampleCode.zip
      SampleCode.hkl - data corrected using  SADABS
      SampleCode_merge_none.hkl - equivalents and friedels kept separate  (See "Final Data Set" table in this report)
      SampleCode.p4p - skeleton p4p file for use with XPREP
      SampleCode.htm - this summary file


Southampton CIF entry fields

_exptl_absorpt_process_details       'SADABS 2007/2 (Sheldrick, G.M., 2007)'

_diffrn_ambient_temperature           120(2)
_diffrn_radiation_wavelength           0.71073
_diffrn_radiation_type                      MoK\a
_diffrn_radiation_source                  'Bruker-Nonius FR591 rotating anode'
_diffrn_radiation_monochromator    'graphite'
_diffrn_measurement_device_type   'Bruker-Nonius Roper CCD camera on \k-goniostat'
_diffrn_measurement_method          '\f & \w scans'
_diffrn_detector_area_resol_mean   '9.091'

_computing_data_collection        
'COLLECT (Hooft, R.W.W., 1998)'
_computing_cell_refinement        
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)'
#Although determined using DirAx, the cell is refined in the HKL
#package during data reduction
_computing_data_reduction         
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)' 

Software References

Unit Cell Determination using DirAx:
Duisenberg AJM. Indexing in Single-Crystal Diffractometry with an Obstinate List of Reflections. Journal of Applied Crystallography 1992; 25: 92-96.
Duisenberg AJM, Hooft RWW, Schreurs AMM, Kroon J. Accurate cells from area-detector images. Journal of Applied Crystallography 2000; 33: 893-898.

Diffractometer control and strategy calculation using COLLECT:
Hooft RWW. COLLECT data collection software, Nonius B.V., 1998.

Standard Data Reduction using HKL(Denzo & Scalepack):
Otwinowski Z, Minor W. Processing of X-ray diffraction data collected in oscillation mode. Macromolecular Crystallography, Pt A. 1997; 276: 307-326.

Absorption Correction
Sheldrick, G. M. (2007). SADABS. Version 2007/2. Bruker AXS Inc., Madison, Wisconsin, USA.

Non-Standard Data Reduction using EvalCCD:
Duisenberg AJM, Kroon-Batenburg LMJ, Schreurs AMM. An intensity evaluation method: EVAL-14. Journal of Applied Crystallography 2003; 36: 220-229.