EPSRC National Crystallography Service

Data Collection Summary

Summary report for Directory: diska/01src497

Report generated May 29, 2001; 08:22:21

Unit cell

80010 reflections with 1.00°<theta<26.02° (resolution between 20.40A and 0.81A) were used for unit cell refinement

Symmetry used
in scalepack
p2
a (Angstrom)24.7998 +/- 0.0003
b (Angstrom)16.6806 +/- 0.0002
c (Angstrom)24.9057 +/- 0.0003
alpha (°) 90.000
beta (°)107.1610 +/- 0.0010
gamma (°) 90.000
Volume (A**3)9844.2 +/- 0.2
Mosaicity (°)0.4620 +/- 0.0010

Data collection

Summary

Total number of images collected504
Total exposure time335.3 minutes
Data collection exposure time331.1 minutes
Data collection wall-clock time393.3 minutes

Experimental Conditions

Wavelength0.71073 A
Generator setting50kV 85mA

Scans

TypeName# imagesTotal
Rotation
Per frame
Rotation
Exposure
per frame
Detector
distance
Used in
scaling
cell determinationi01f10 10.0° phi1.000°25 seconds60.00 mmNo
data collections01f362362.0° phi1.000°40 seconds45.00 mmYes
data collections02f64 64.0° omega1.000°40 seconds45.00 mmYes
data collections03f68 68.0° omega1.000°40 seconds45.00 mmYes

Scalepack Scaling

Deleted observations

Zero sigma or profile test  48
Overload or incomplete profile2780
Sigma cutoff 173

Final Data Set

Scale factor10.00
Number of 'full' reflections 15249
Number of 'partial' reflections156078
Total number of integrated reflections 97975
Total number of unique reflections 20048
Data completeness  99.4%
Resolution range20.40-0.81 A
Theta range1.00°-26.02°
Average Intensity   10.0
Average Sigma(I)    1.3
Overall R-merge (linear)  0.175

Crystal Information

Collection Temperature   120 K
Crystal Size    x   x  
Crystal Colour  
Crystal Shape  

SORTAV Absorption correction Summary

R(merge) on F2
Before
After
R
0.199
0.190
Rw
4.099
4.177
Max Transmission Factor
0.713E+00
Min Transmission Factor
0.713E+00

N.B. The scaling summary is redundant as outliers are treated during the absorption correction using SORTAV. The quoted data completeness is for the stated resolution ranges, and the Overall R-merge is that before the absorption correction. The SORTAV transmission factors are based on a crude approximation and the expected formula (not always correct!). For more information visit the service web site at: http://www.soton.ac.uk/~xservice/