Data Collection Summary kccd2
(damien) |
Summary report for Directory: diska/2009src0216r
Report generated Mar 02, 2009; 18:05:44
Unit cell
16828 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement
Symmetry used in scalepack | p2 |
a (Angstrom) | 6.2066 +/- 0.0003 |
b (Angstrom) | 17.1960 +/- 0.0008 |
c (Angstrom) | 15.1712 +/- 0.0006 |
alpha (°) | 90.000 |
beta (°) | 90.442 +/- 0.003 |
gamma (°) | 90.000 |
Volume (A**3) | 1619.15 +/- 0.13 |
---|
Mosaicity (°) | 0.626 +/- 0.002 |
Data collection
Summary
Total number of images collected | 251 |
Total exposure time | 149.1 minutes |
Data collection exposure time | 147.1 minutes |
Data collection wall-clock time | 160.1 minutes |
Experimental Conditions
Wavelength | 0.71073 A |
Crystal to detector distance | 32.00 mm |
Scans
Type | Name | # images | Total Rotation | Per frame Rotation | Exposure per frame | Used in scaling |
data collection | s01f | 123 | 246.0° phi | 2.000° | 36 seconds | Yes |
data collection | s02f | 13 | 26.0° omega | 2.000° | 36 seconds | Yes |
data collection | s03f | 25 | 50.0° omega | 2.000° | 36 seconds | Yes |
data collection | s04f | 31 | 62.0° omega | 2.000° | 36 seconds | Yes |
data collection | s05f | 51 | 102.0° omega | 2.000° | 36 seconds | Yes |
Phi/Chi | i01f - i08f | 8 | | | 15 seconds |
Scalepack Scaling
Deleted observations
Rejected | 5 |
Zero sigma or profile test | 7 |
Overload or incomplete profile | 500 |
Sigma cutoff | 446 |
High resolution limit | 72 |
Final Data Set
Scale factor | 10.00 |
Number of 'full' reflections | 13323 |
Number of 'partial' reflections | 10606 |
Total number of integrated reflections | 22406 |
Total number of unique reflections | 3822 |
Data completeness | 99.9% |
Resolution range | 7.00-0.77 A |
Theta range | 2.91°-27.48° |
Average Intensity | 13.1 |
Average Sigma(I) | 0.9 |
Overall R-merge (linear) | 0.206 |
Sadabs Results
Parameter refinement on 13981 reflections reduced R(int) from 0.1354 to 0.0937
Before rejection, 22341 reflections total and 3847 unique
After rejection, 18947 reflections total and 3828 unique
Runs
Run 2-theta R(int) Incid. factors Diffr. factors K Total I>2sig(I)
1 12.6 0.1294 0.724 - 0.900 0.967 - 1.222 2.045 9819 5620
2 12.0 0.1308 0.583 - 0.608 0.967 - 1.098 2.049 899 520
3 12.0 0.1197 0.750 - 0.849 1.014 - 1.159 2.049 1899 1128
4 12.0 0.1195 0.732 - 0.886 0.987 - 1.152 1.990 2373 1450
5 12.0 0.1183 0.756 - 0.943 1.014 - 1.215 1.962 3957 2374
Ratio of minimum to maximum apparent transmission: 0.791249
Metadata
  Group   | Service   |
  Operator   | Sam Callear   |
  Sample Owner   | Dr J Spencer   |
  Local Code   | RRPSAN02   |
  Formula   | C14H21ClN3PPdS   |
  Crystal Colour   | Colourless   |
  Crystal Habit   | Plate   |
  Crystal Size   | 0.09 x 0.07 x 0.01 (mm3)  |
  Temperature   | 120(K)  |
  Generator   | 50 (kV)   85 (mA)  |
  Primary Solvent   |   |
  Other Solvents   |   |
File Explanations
SampleCode.zip
SampleCode.hkl
- data corrected using SADABS
SampleCode_merge_none.hkl
- equivalents and friedels kept separate (See "Final Data
Set" table in this report)
SampleCode.p4p
- skeleton p4p file for use with XPREP
SampleCode.htm
- this summary file
Southampton CIF entry fields
_exptl_absorpt_process_details 'SADABS 2007/2
(Sheldrick, G.M.,
2007)'
_diffrn_ambient_temperature
120(2)
_diffrn_radiation_wavelength
0.71073
_diffrn_radiation_type
MoK\a
_diffrn_radiation_source
'Bruker-Nonius FR591 rotating anode'
_diffrn_radiation_monochromator '10cm confocal mirrors'
_diffrn_measurement_device_type 'Bruker-Nonius APEX II CCD camera on \k-goniostat'
_diffrn_measurement_method
'\f & \w scans'
_diffrn_detector_area_resol_mean '4096x4096pixels / 62x62mm'
_computing_data_collection
'COLLECT (Hooft, R.W.W., 1998)'
_computing_cell_refinement
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)'
#Although determined using DirAx, the cell is refined in the HKL
#package during data reduction
_computing_data_reduction
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W.,
1998)'
Software References
Unit Cell Determination using DirAx:
Duisenberg AJM. Indexing in Single-Crystal Diffractometry with an
Obstinate List of Reflections. Journal
of Applied Crystallography 1992; 25: 92-96.
Duisenberg AJM, Hooft RWW, Schreurs AMM, Kroon J. Accurate cells from
area-detector images. Journal of
Applied Crystallography 2000; 33:
893-898.
Diffractometer control and strategy
calculation using COLLECT:
Hooft RWW.
COLLECT data collection software, Nonius B.V., 1998.
Standard
Data Reduction using HKL(Denzo
& Scalepack):
Otwinowski Z, Minor W.
Processing of X-ray diffraction data collected in oscillation mode. Macromolecular Crystallography, Pt
A. 1997; 276: 307-326.
Absorption Correction
Sheldrick, G. M. (2007). SADABS. Version 2007/2. Bruker AXS Inc., Madison,
Wisconsin, USA.
Non-Standard Data Reduction using
EvalCCD:
Duisenberg AJM, Kroon-Batenburg LMJ, Schreurs AMM. An intensity
evaluation method: EVAL-14. Journal
of Applied Crystallography 2003; 36: 220-229.
For more information visit the service web site at: http://www.soton.ac.uk/~xservice/