EPSRC National Crystallography Service
Data Collection Summary kccd1 (dellboy)


Summary report for Directory: diska/2007src0011

Report generated Jan 16, 2007; 16:55:40

Unit cell

25190 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement

Symmetry used
in scalepack
p4
a (Angstrom)11.49110 +/- 0.00010
b (Angstrom)11.49110 +/- 0.00010
c (Angstrom)16.7887 +/- 0.0002
alpha (°) 90.000
beta (°) 90.000
gamma (°) 90.000
Volume (A**3)2216.87 +/- 0.04
Mosaicity (°)0.4330 +/- 0.0010

Data collection

Summary

Total number of images collected278
Total exposure time46.5 minutes
Data collection exposure time45.9 minutes
Data collection wall-clock time73.9 minutes

Experimental Conditions

Wavelength0.71073 A
Crystal to detector distance30.00 mm

Scans

TypeName# imagesTotal
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collections01f181362.0° phi2.000°10 secondsYes
data collections02f17 34.0° omega2.000°10 secondsYes
data collections03f17 34.0° omega2.000°10 secondsYes
data collections04f55110.0° omega2.000°10 secondsYes
Phi/Chii01f - i08f84 seconds

Scalepack Scaling

Deleted observations

Rejected 273
Zero sigma or profile test   2
Overload or incomplete profile1402
Sigma cutoff  26
High resolution limit   9

Final Data Set

Scale factor range9.00-10.39
Number of 'full' reflections 30119
Number of 'partial' reflections 13829
Total number of integrated reflections 40563
Total number of unique reflections  2629
Data completeness  99.9%
Resolution range7.00-0.77 A
Theta range2.91°-27.48°
Average Intensity  371.3
Average Sigma(I)    4.1
Overall R-merge (linear)  0.033

Sadabs Results

Parameter refinement on 40557 reflections reduced R(int) from 0.0818 to 0.0331

Before rejection, 42256 reflections total and 2637 unique

After rejection, 41685 reflections total and 2636 unique

Runs

  Run 2-theta  R(int)  Incid. factors  Diffr. factors    K     Total I>2sig(I)
    1    7.4  0.0325   0.886 - 1.030   0.974 - 1.138   1.606   28318   27163
    2    6.8  0.0303   0.941 - 1.050   0.974 - 1.087   1.575    2438    2346
    3    6.8  0.0302   0.783 - 0.824   0.966 - 1.073   1.578    2395    2304
    4    6.8  0.0309   0.891 - 0.951   0.966 - 1.103   1.598    8534    8191
Ratio of minimum to maximum apparent transmission: 0.848413

Metadata

  Group    Service  
  Operator   Simon Coles  
  Sample Owner   Davies  
  Local Code   306-725-1-A  
  Formula   C18H24N5O2S2P3  
  Crystal Colour    Colourless  
  Crystal Habit    Prism  
  Crystal Size   0.45 x 0.45 x 0.4 (mm3) 
  Temperature   120(K) 
  Generator   50 (kV)   85 (mA) 
  Primary Solvent     
  Other Solvents     

File Explanations

SampleCode.zip
      SampleCode.hkl - data corrected using  SADABS
      SampleCode_merge_none.hkl - equivalents and friedels kept separate  (See "Final Data Set" table in this report)
      SampleCode.p4p - skeleton p4p file for use with XPREP
      SampleCode.htm - this summary file


Southampton CIF entry fields

_exptl_absorpt_process_details       'SADABS V2.10 (Sheldrick, G.M., 2003)'

_diffrn_ambient_temperature           120(2)
_diffrn_radiation_wavelength           0.71073
_diffrn_radiation_type                      MoK\a
_diffrn_radiation_source                  'Bruker-Nonius FR591 rotating anode'
_diffrn_radiation_monochromator    'graphite'
_diffrn_measurement_device_type  'Bruker-Nonius 95mm CCD camera on \k-goniostat'
_diffrn_measurement_method          '\f & \w scans'
_diffrn_detector_area_resol_mean   9.091

_computing_data_collection        
'COLLECT (Hooft, R.W.W., 1998)'
_computing_cell_refinement        
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)'
#Although determined using DirAx, the cell is refined in the HKL
#package during data reduction
_computing_data_reduction         
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)' 

Software References

Unit Cell Determination using DirAx:
Duisenberg AJM. Indexing in Single-Crystal Diffractometry with an Obstinate List of Reflections. Journal of Applied Crystallography 1992; 25: 92-96.
Duisenberg AJM, Hooft RWW, Schreurs AMM, Kroon J. Accurate cells from area-detector images. Journal of Applied Crystallography 2000; 33: 893-898.

Diffractometer control and strategy calculation using COLLECT:
Hooft RWW. COLLECT data collection software, Nonius B.V., 1998.

Standard Data Reduction using HKL(Denzo & Scalepack):
Otwinowski Z, Minor W. Processing of X-ray diffraction data collected in oscillation mode. Macromolecular Crystallography, Pt A. 1997; 276: 307-326.

Absorption Correction
Sheldrick, G. M. (2003). SADABS. Version 2.10. Bruker AXS Inc., Madison, Wisconsin, USA.

Non-Standard Data Reduction using EvalCCD:
Duisenberg AJM, Kroon-Batenburg LMJ, Schreurs AMM. An intensity evaluation method: EVAL-14. Journal of Applied Crystallography 2003; 36: 220-229.

For more information visit the service web site at: http://www.soton.ac.uk/~xservice/