EPSRC National Crystallography Service
Data Collection Summary kccd1 (dellboy)


Summary report for Directory: diska/2008lsh001

Report generated Jun 23, 2008; 18:18:22

Unit cell

13268 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement

Symmetry used
in scalepack
p2
a (Angstrom)9.6630 +/- 0.0003
b (Angstrom)9.2888 +/- 0.0002
c (Angstrom)19.8221 +/- 0.0005
alpha (°) 90.000
beta (°)90.0315 +/- 0.0014
gamma (°) 90.000
Volume (A**3)1779.19 +/- 0.08
Mosaicity (°)0.424 +/- 0.002

Data collection

Summary

Total number of images collected221
Total exposure time55.4 minutes
Data collection exposure time54.0 minutes
Data collection wall-clock time72.5 minutes

Experimental Conditions

Wavelength0.71073 A
Crystal to detector distance30.00 mm

Scans

TypeName# imagesTotal
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collections01f109163.5° phi1.500°15 secondsYes
data collections02f36 54.0° omega1.500°15 secondsYes
data collections03f68102.0° omega1.500°15 secondsYes
Phi/Chii01f - i08f810 seconds

Scalepack Scaling

Deleted observations

Rejected  24
Zero sigma or profile test   3
Overload or incomplete profile 696
Sigma cutoff  89
High resolution limit  17

Final Data Set

Scale factor10.00
Number of 'full' reflections 12810
Number of 'partial' reflections  8655
Total number of integrated reflections 20634
Total number of unique reflections  4260
Data completeness  98.1%
Resolution range7.00-0.77 A
Theta range2.91°-27.48°
Average Intensity   78.1
Average Sigma(I)    2.6
Overall R-merge (linear)  0.069

Sadabs Results

Estimated minimum and maximum transmission: 0.6277 0.7456

Metadata

  Group      
  Operator   Susanne Huth  
  Sample Owner     
  Local Code   PAA/CF3  
  Formula   C9 H8 F3 N1 O1  
  Crystal Colour    Colourless  
  Crystal Habit    Needle  
  Crystal Size   0.5 x 0.16 x 0.16 (mm3) 
  Temperature   120(K) 
  Generator   50 (kV)   85 (mA) 
  Primary Solvent     
  Other Solvents     

File Explanations

SampleCode.zip
      SampleCode.hkl - data corrected using  SADABS
      SampleCode_merge_none.hkl - equivalents and friedels kept separate  (See "Final Data Set" table in this report)
      SampleCode.p4p - skeleton p4p file for use with XPREP
      SampleCode.htm - this summary file


Southampton CIF entry fields

_exptl_absorpt_process_details       'SADABS 2007/2 (Sheldrick, G.M., 2007)'

_diffrn_ambient_temperature           120(2)
_diffrn_radiation_wavelength           0.71073
_diffrn_radiation_type                      MoK\a
_diffrn_radiation_source                  'Bruker-Nonius FR591 rotating anode'
_diffrn_radiation_monochromator    'graphite'
_diffrn_measurement_device_type   'Bruker-Nonius Roper CCD camera on \k-goniostat'
_diffrn_measurement_method          '\f & \w scans'
_diffrn_detector_area_resol_mean   '9.091'

_computing_data_collection        
'COLLECT (Hooft, R.W.W., 1998)'
_computing_cell_refinement        
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)'
#Although determined using DirAx, the cell is refined in the HKL
#package during data reduction
_computing_data_reduction         
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)' 

Software References

Unit Cell Determination using DirAx:
Duisenberg AJM. Indexing in Single-Crystal Diffractometry with an Obstinate List of Reflections. Journal of Applied Crystallography 1992; 25: 92-96.
Duisenberg AJM, Hooft RWW, Schreurs AMM, Kroon J. Accurate cells from area-detector images. Journal of Applied Crystallography 2000; 33: 893-898.

Diffractometer control and strategy calculation using COLLECT:
Hooft RWW. COLLECT data collection software, Nonius B.V., 1998.

Standard Data Reduction using HKL(Denzo & Scalepack):
Otwinowski Z, Minor W. Processing of X-ray diffraction data collected in oscillation mode. Macromolecular Crystallography, Pt A. 1997; 276: 307-326.

Absorption Correction
Sheldrick, G. M. (2007). SADABS. Version 2007/2. Bruker AXS Inc., Madison, Wisconsin, USA.

Non-Standard Data Reduction using EvalCCD:
Duisenberg AJM, Kroon-Batenburg LMJ, Schreurs AMM. An intensity evaluation method: EVAL-14. Journal of Applied Crystallography 2003; 36: 220-229.

For more information visit the service web site at: http://www.soton.ac.uk/~xservice/