Unit cell

20829 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement

Symmetry used
in scalepack
p1
a (Angstrom)5.7366 +/- 0.0002
b (Angstrom)8.5492 +/- 0.0005
c (Angstrom)10.6373 +/- 0.0006
alpha (°)93.974 +/- 0.003
beta (°)99.351 +/- 0.003
gamma (°)95.314 +/- 0.003
Volume (A**3)510.65 +/- 0.05
Mosaicity (°)1.293 +/- 0.003

Data collection

Summary

Total number of images collected349
Total exposure time116.3 minutes
Data collection exposure time114.9 minutes
Data collection wall-clock time152.1 minutes

Experimental Conditions

Wavelength0.71073 A
Crystal to detector distance30.00 mm

Scans

TypeName# imagesTotal
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collections01f179358.0° phi2.000°20 secondsYes
data collections02f17 34.0° omega2.000°20 secondsYes
data collections03f16 32.0° omega2.000°20 secondsYes
data collections04f55110.0° omega2.000°20 secondsYes
data collections05f23 46.0° omega2.000°20 secondsYes
data collections06f51102.0° omega2.000°20 secondsYes
Phi/Chii01f - i08f810 seconds

Scalepack Scaling

Deleted observations

Rejected  16
Zero sigma or profile test   1
Overload or incomplete profile 540
Sigma cutoff  33
High resolution limit  14

Final Data Set

Scale factor10.00
Number of 'full' reflections  2661
Number of 'partial' reflections 10485
Total number of integrated reflections 12304
Total number of unique reflections  2338
Data completeness  99.7%
Resolution range7.00-0.77 A
Theta range2.91°-27.48°
Average Intensity  320.4
Average Sigma(I)    6.6
Overall R-merge (linear)  0.045

Sadabs Results

Estimated minimum and maximum transmission: 0.6408 0.7456

Metadata

  Group    Mike Hursthouse  
  Operator   Susanne Huth  
  Sample Owner     
  Local Code   MAA/OCOEt  
  Formula   C11 H13 N1 O3  
  Crystal Colour    Colourless  
  Crystal Habit   Cut Block  
  Crystal Size   0.6 x 0.5 x 0.2 (mm3) 
  Temperature   120(K) 
  Generator   50 (kV)   85 (mA) 
  Primary Solvent     
  Other Solvents     

File Explanations

SampleCode.zip
      SampleCode.hkl - data corrected using  SADABS
      SampleCode_merge_none.hkl - equivalents and friedels kept separate  (See "Final Data Set" table in this report)
      SampleCode.p4p - skeleton p4p file for use with XPREP
      SampleCode.htm - this summary file


Southampton CIF entry fields

_exptl_absorpt_process_details       'SADABS 2007/2 (Sheldrick, G.M., 2007)'

_diffrn_ambient_temperature           120(2)
_diffrn_radiation_wavelength           0.71073
_diffrn_radiation_type                      MoK\a
_diffrn_radiation_source                  'Bruker-Nonius FR591 rotating anode'
_diffrn_radiation_monochromator    'graphite'
_diffrn_measurement_device_type   'Bruker-Nonius Roper CCD camera on \k-goniostat'
_diffrn_measurement_method          '\f & \w scans'
_diffrn_detector_area_resol_mean   '9.091'

_computing_data_collection        
'COLLECT (Hooft, R.W.W., 1998)'
_computing_cell_refinement        
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)'
#Although determined using DirAx, the cell is refined in the HKL
#package during data reduction
_computing_data_reduction         
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)' 

Software References

Unit Cell Determination using DirAx:
Duisenberg AJM. Indexing in Single-Crystal Diffractometry with an Obstinate List of Reflections. Journal of Applied Crystallography 1992; 25: 92-96.
Duisenberg AJM, Hooft RWW, Schreurs AMM, Kroon J. Accurate cells from area-detector images. Journal of Applied Crystallography 2000; 33: 893-898.

Diffractometer control and strategy calculation using COLLECT:
Hooft RWW. COLLECT data collection software, Nonius B.V., 1998.

Standard Data Reduction using HKL(Denzo & Scalepack):
Otwinowski Z, Minor W. Processing of X-ray diffraction data collected in oscillation mode. Macromolecular Crystallography, Pt A. 1997; 276: 307-326.

Absorption Correction
Sheldrick, G. M. (2007). SADABS. Version 2007/2. Bruker AXS Inc., Madison, Wisconsin, USA.

Non-Standard Data Reduction using EvalCCD:
Duisenberg AJM, Kroon-Batenburg LMJ, Schreurs AMM. An intensity evaluation method: EVAL-14. Journal of Applied Crystallography 2003; 36: 220-229.

For more information visit the service web site at: http://www.soton.ac.uk/~xservice/