EPSRC National Crystallography Service
Data Collection Summary kccd1 (dellboy)


Summary report for Directory: diska/2010src0795

Report generated Aug 23, 2010; 10:13:44

Unit cell

29464 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement

Symmetry used
in scalepack
c2
a (Angstrom)19.6553 +/- 0.0012
b (Angstrom)4.0406 +/- 0.0002
c (Angstrom)32.653 +/- 0.002
alpha (°) 90.000
beta (°)101.378 +/- 0.002
gamma (°) 90.000
Volume (A**3)2542.3 +/- 0.3
Mosaicity (°)1.102 +/- 0.002

Data collection

Summary

Total number of images collected757
Total exposure time11.4 hours
Data collection exposure time11.3 hours
Data collection wall-clock time12.3 hours

Experimental Conditions

Wavelength0.71073 A
Crystal to detector distance48.00 mm

Scans

TypeName# imagesTotal
Rotation
Per frame
Rotation
Exposure
per frame
Used in
scaling
data collections01f330297.0° phi0.900°54 secondsYes
data collections02f157141.3° omega0.900°54 secondsYes
data collections03f105 94.5° omega0.900°54 secondsYes
data collections04f157141.3° omega0.900°54 secondsYes
Phi/Chii01f - i08f840 seconds

Scalepack Scaling

Deleted observations

Rejected  14
Zero sigma or profile test  42
Overload or incomplete profile 554
Sigma cutoff  39
High resolution limit  19

Final Data Set

Scale factor10.00
Number of 'full' reflections     0
Number of 'partial' reflections 13570
Total number of integrated reflections 10033
Total number of unique reflections  3030
Data completeness  91.1%
Resolution range7.00-0.77 A
Theta range2.91°-27.48°
Average Intensity   92.5
Average Sigma(I)    4.0
Overall R-merge (linear)  0.064

Sadabs Results

Estimated minimum and maximum transmission: 0.4716 0.7456

Metadata

  Group    Service  
  Operator   Graham Tizzard  
  Sample Owner   Dr J Spencer  
  Local Code   GRE.Js4  
  Formula   C15H10Cl1N1O1  
  Crystal Colour    Yellow  
  Crystal Habit    Rod  
  Crystal Size   0.45 x 0.07 x 0.05 (mm3) 
  Temperature   120(K) 
  Generator   50 (kV)   85 (mA) 
  Primary Solvent     
  Other Solvents     

File Explanations

SampleCode.zip
      SampleCode.hkl - data corrected using  SADABS
      SampleCode_merge_none.hkl - equivalents and friedels kept separate  (See "Final Data Set" table in this report)
      SampleCode.p4p - skeleton p4p file for use with XPREP
      SampleCode.htm - this summary file


Southampton CIF entry fields

_exptl_absorpt_process_details       'SADABS 2007/2 (Sheldrick, G.M., 2007)'

_diffrn_ambient_temperature           120(2)
_diffrn_radiation_wavelength           0.71073
_diffrn_radiation_type                      MoK\a
_diffrn_radiation_source                  'Bruker-Nonius FR591 rotating anode'
_diffrn_radiation_monochromator    'graphite'
_diffrn_measurement_device_type   'Bruker-Nonius Roper CCD camera on \k-goniostat'
_diffrn_measurement_method          '\f & \w scans'
_diffrn_detector_area_resol_mean   '9.091'

_computing_data_collection        
'COLLECT (Hooft, R.W.W., 1998)'
_computing_cell_refinement        
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)'
#Although determined using DirAx, the cell is refined in the HKL
#package during data reduction
_computing_data_reduction         
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)' 

Software References

Unit Cell Determination using DirAx:
Duisenberg AJM. Indexing in Single-Crystal Diffractometry with an Obstinate List of Reflections. Journal of Applied Crystallography 1992; 25: 92-96.
Duisenberg AJM, Hooft RWW, Schreurs AMM, Kroon J. Accurate cells from area-detector images. Journal of Applied Crystallography 2000; 33: 893-898.

Diffractometer control and strategy calculation using COLLECT:
Hooft RWW. COLLECT data collection software, Nonius B.V., 1998.

Standard Data Reduction using HKL(Denzo & Scalepack):
Otwinowski Z, Minor W. Processing of X-ray diffraction data collected in oscillation mode. Macromolecular Crystallography, Pt A. 1997; 276: 307-326.

Absorption Correction
Sheldrick, G. M. (2007). SADABS. Version 2007/2. Bruker AXS Inc., Madison, Wisconsin, USA.

Non-Standard Data Reduction using EvalCCD:
Duisenberg AJM, Kroon-Batenburg LMJ, Schreurs AMM. An intensity evaluation method: EVAL-14. Journal of Applied Crystallography 2003; 36: 220-229.

For more information visit the service web site at: http://www.soton.ac.uk/~xservice/