EPSRC
National Crystallography Service |
Data Collection Summary kccd1
(dellboy) |
Summary report for Directory: diska/2005src1568
Report generated Dec 01, 2005; 13:23:16
Unit cell
9805 reflections with 2.91°<theta<27.48° (resolution between 7.00A and 0.77A) were used for unit cell refinement
Symmetry used in scalepack | p2 |
a (Angstrom) | 11.5026 +/- 0.0003 |
b (Angstrom) | 13.7030 +/- 0.0002 |
c (Angstrom) | 27.1789 +/- 0.0007 |
alpha (°) | 90.000 |
beta (°) | 100.9996 +/- 0.0008 |
gamma (°) | 90.000 |
Volume (A**3) | 4205.24 +/- 0.17 |
---|
Mosaicity (°) | 0.5270 +/- 0.0010 |
Data collection
Summary
Total number of images collected | 538 |
Total exposure time | 91.3 minutes |
Data collection exposure time | 90.2 minutes |
Data collection wall-clock time | 138.8 minutes |
Experimental Conditions
Wavelength | 0.71073 A |
Crystal to detector distance | 40.00 mm |
Scans
Type | Name | # images | Total Rotation | Per frame Rotation | Exposure per frame | Used in scaling |
data collection | s01f | 308 | 308.0° phi | 1.000° | 10 seconds | Yes |
data collection | s02f | 34 | 34.0° omega | 1.000° | 10 seconds | Yes |
data collection | s03f | 64 | 64.0° omega | 1.000° | 10 seconds | Yes |
data collection | s04f | 124 | 124.0° omega | 1.000° | 10 seconds | Yes |
Phi/Chi | i01f - i08f | 8 | | | 8 seconds |
Scalepack Scaling
Deleted observations
Rejected | 233 |
Zero sigma or profile test | 7 |
Overload or incomplete profile | 1592 |
Sigma cutoff | 48 |
High resolution limit | 105 |
Final Data Set
Scale factor range | 8.65-11.26 |
Number of 'full' reflections | 18357 |
Number of 'partial' reflections | 37622 |
Total number of integrated reflections | 51477 |
Total number of unique reflections | 10001 |
Data completeness | 99.8% |
Resolution range | 7.00-0.77 A |
Theta range | 2.91°-27.48° |
Average Intensity | 21.3 |
Average Sigma(I) | 1.4 |
Overall R-merge (linear) | 0.066 |
Sadabs Results
Parameter refinement on 28216 reflections reduced R(int) from 0.0854 to 0.0462
Before rejection, 54460 reflections total and 10045 unique
After rejection, 53933 reflections total and 10041 unique
Runs
Run 2-theta R(int) Incid. factors Diffr. factors K Total I>2sig(I)
1 15.6 0.0661 0.907 - 1.140 0.887 - 1.089 1.213 31226 17856
2 -15.0 0.0629 0.668 - 0.710 0.888 - 1.187 1.259 3305 1937
3 -15.0 0.0715 0.684 - 0.773 0.920 - 1.062 1.258 6446 3773
4 -15.0 0.0701 0.899 - 1.082 0.916 - 1.094 1.209 12956 7280
Ratio of minimum to maximum apparent transmission: 0.746863
Metadata
  Group   | Service   |
  Operator   | Simon Coles   |
  Sample Owner   | Greeves   |
  Local Code   | 384/1   |
  Formula   | C24H32O   |
  Crystal Colour   | Colourless   |
  Crystal Habit   | Block   |
  Crystal Size   | 0.4 x 0.4 x 0.3 (mm3)  |
  Temperature   | 120(K)  |
  Generator   | 50 (kV)   85 (mA)  |
  Primary Solvent   |   |
  Other Solvents   |   |
File Explanations
SampleCode.zip
SampleCode.hkl
- data corrected using SADABS
SampleCode_merge_none.hkl
- equivalents and friedels kept separate (See "Final Data
Set" table in this report)
SampleCode.p4p
- skeleton p4p file for use with XPREP
SampleCode.htm
- this summary file
Southampton CIF entry fields
_exptl_absorpt_process_details 'SADABS V2.10
(Sheldrick, G.M.,
2003)'
_diffrn_ambient_temperature
120(2)
_diffrn_radiation_wavelength
0.71073
_diffrn_radiation_type
MoK\a
_diffrn_radiation_source
'Bruker-Nonius FR591 rotating anode'
_diffrn_radiation_monochromator 'graphite'
_diffrn_measurement_device_type 'Bruker-Nonius 95mm CCD
camera on \k-goniostat'
_diffrn_measurement_method
'\f & \w scans'
_diffrn_detector_area_resol_mean 9.091
_computing_data_collection
'COLLECT (Hooft, R.W.W., 1998)'
_computing_cell_refinement
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W., 1998)'
#Although determined using DirAx, the cell is refined in the HKL
#package during data reduction
_computing_data_reduction
'DENZO (Otwinowski & Minor, 1997) & COLLECT (Hooft, R.W.W.,
1998)'
Software References
Unit Cell Determination using DirAx:
Duisenberg AJM. Indexing in Single-Crystal Diffractometry with an
Obstinate List of Reflections. Journal
of Applied Crystallography 1992; 25: 92-96.
Duisenberg AJM, Hooft RWW, Schreurs AMM, Kroon J. Accurate cells from
area-detector images. Journal of
Applied Crystallography 2000; 33:
893-898.
Diffractometer control and strategy
calculation using COLLECT:
Hooft RWW.
COLLECT data collection software, Nonius B.V., 1998.
Standard
Data Reduction using HKL(Denzo
& Scalepack):
Otwinowski Z, Minor W.
Processing of X-ray diffraction data collected in oscillation mode. Macromolecular Crystallography, Pt
A. 1997; 276: 307-326.
Absorption Correction
Sheldrick, G. M. (2003). SADABS. Version 2.10. Bruker AXS Inc., Madison,
Wisconsin, USA.
Non-Standard Data Reduction using
EvalCCD:
Duisenberg AJM, Kroon-Batenburg LMJ, Schreurs AMM. An intensity
evaluation method: EVAL-14. Journal
of Applied Crystallography 2003; 36: 220-229.
For more information visit the service web site at: http://www.soton.ac.uk/~xservice/